Skip to content
Logo
Model 6403 90cm Helmholtz coil system has twin coils and is normally used as a magnetic field immunity test system. A magnetic field is produced by two coils that is essentially homogenous throughout the volume between the face of the coils and one-third diameter of the coils at it’s centre. The coils are supported by a non-metallic framework. A rotating, height-adjustable pedestal is capable of holding articles weighing up to 34kg. The Coil has been initially calibrated at the NPL in terms of uT/amp at 50Hz. For the end user’s convenience NPL have also quoted a figure of uTesla per millivolt as measured across an installed shunt resistor with measuring terminal placed at the base of the Helmholtz. This was done at 60Hz and 2kHz. Built in on the base is the feed connector.