| Measuring range |
0.5 to 600 V/m |
| Overload indicator on |
> 1000 V/m |
| Maximum field level |
1500 V/m |
| Frequency range |
10 MHz to 6 GHz |
| Frequency response (with correction factors applied)1 |
± 0.6 dB @ 30 MHz |
| Frequency response (uncorrected) |
10 MHz to 1 GHz @ ± 1.5 dB |
|
1 GHz to 6 GHz @ ± 3.0 dB |
| Resolution |
0.01 V/m |
| Linearity |
± 0.5 dB +/- 0.5 V/m |
| Isotropic deviation |
< ± 0.25 dB @ 1 GHz |
| Measurement speed (X,Y, Z & ETot) |
60 samples/s |
| Shape |
Spherical |
| Weight |
50 g (1.77oz) |
| Electrical measuring volume |
52 cm3 |
| Antenna elements |
1.2 cm monopole |
| Spherical housing diameter |
2.5 cm (0.98 in) |
| Sensor protection caps |
1.4 cm (height), 1.1 cm (diameter) |
| Operating temperature range |
15 °C to 35 °C (59 °F to 95 °F) @ 10% to 90% RH (non-condensing) |
| Calibration data |
ISO17025 accredited calibration (optional) |
| Optical LASER power |
0.5 Watt at aperture at 808 nm |
| F.O. connector LASER |
FC/PC 200/230 µm fibre, 1.5 m fixed and 10 m extension2 |
| F.O. connector data |
FSMA 200/230µm fibre, 1.5m fixed and 10 m extension2 |